Appendix B: Error Codes
General Errors,
Cont’d
Code
Cause
2100
FLUID_ERR
No fluid detected during a dispense, prime, or
maintenance cycle in a program.
Probable Causes:
•
Ball used to block the sensor is missing or stuck
below the infrared detector.
•
Infrared detector or fluid sensor is no longer
functioning.
•
Dispense valve has not closed and fluid is still
flowing.
•
Air bubble present in the detection path.
2200
FLOW_ERR
No flow detected – occurs at the end of a dispense, if
no fluid flow was detected during the dispense
operation.
Probable Causes:
•
Fluid or sensor housing is opaque, not allowing
the infrared detector to toggle.
•
Infrared detector or fluid sensor failed.
•
Main PCB failed to detect the fluid sensor.
2300
INVALID_PLATE_TYPE
Plate type not valid for hardware – occurs at the start
of a program.
Probable Causes:
•
Invalid plate type entered.
•
Protocol is corrupt.
2400
INVALID_LINK_ERR
Failure to get a program to view/run – occurs at
program start time, at program validation time (i.e.,
when programs are saved and/or downloaded), or
when one of the programs within a linked program is
not found or is invalid.
Probable Causes:
•
Memory corruption.
•
Requested program is corrupt or not available.
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...