110
|
Chapter 4: Preventive Maintenance
AUTOPRIME
The AutoPrime feature can be enabled to automatically prime the washer after
it has been idle for a user-specified amount of time. The volume, flow rate, and
soaking parameters specified in the
AUTOPRIME
Prime program are used.
AUTOPRIME
is particularly useful if the washer is used intermittently
throughout the day, to keep the fluid lines and manifold in a wetted condition.
Washers with the external valve module:
AUTOPRIME
uses the
last-used valve (A, B, C, or D) for priming and soaking.
“AUTPRM”
appears only in models with the Ultrasonic Advantage™
(ultrasonic cleaner).
The AutoPrime feature is disabled by default. To enable it:
1. Select
UTIL AUTOPRIME
(or
AUTPRM
)
YES
.
2. Enter the
interval
from
10
to
600
minutes, to indicate how often you
want the
AUTOPRIME
Prime program to run.
3. Press
ENTER
and then return to the Main Menu.
AutoPrime is now enabled and priming will occur each time the washer has
been idle for <
interval
> minutes.
•
To disable the AutoPrime feature, select
UTIL AUTOPRIME
(or
AUTPRM
)
NO.
•
To
halt
the
AUTOPRIME
program, press the
STOP
key.
•
To
view
or
edit
the
AUTOPRIME
program parameters, select
DEFINE
EDIT PRIME
AUTOPRIME
and press the
Options
key to scroll
through the parameters you wish to view or edit. The table below
shows the
AUTOPRIME
program parameters and ranges.
Refer to
Chapter 3
,
Define Prime Program
for more information on Prime
programs and their parameters.
Parameter
Default
Selectable Range
Interval
10 minutes
10 to 600 minutes
Prime Volume
60 ml
1 to 999 ml
Low Flow Prime Volume
(Select CW models only)
0 ml
1 to 999 ml
Prime Flow Rate
7
1-9
Soak After Prime?
Yes
Yes/No
Soak Duration
3 hours, 50 minutes
1 minute to 18 hours
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...