104
|
Chapter 4: Preventive Maintenance
Required Materials
For rinsing/soaking the fluid path, and for cleaning the components:
•
Deionized or distilled water
•
Sodium hypochlorite (NaClO or bleach)
•
70% isopropyl alcohol (or ethanol)
•
Dispense and aspirate tube styluses (supplied with the washer)
•
Phillips head screwdriver
•
9/64” (3.57 mm) hex wrench (supplied with the washer)
•
Lint-free disposable towels
•
Dish soap or other mild cleaner
•
Soft-bristled brush
For replacing o-rings and channel-end seals:
•
O-rings (PN 49941)
•
Channel-end seals (PN 49486)
In the following cleaning procedures, when not otherwise specified,
“water” means use either deionized or distilled water.
Unless otherwise instructed,
always
connect a supply bottle
containing deionized or distilled water to the washer before running
the Maintenance programs.