Detailed Component Descriptions
The MedPlus TotalAlert® Alarm Network can consist of the following: Master
Alarm, Area Alarm, Local Alarm, Area Alarm Network Communications Board,
and gas-specific transducers that include the appropriate DISS, gas-specific,
demand valves.
Master Alarm
The Master Alarm continuously monitors up to 60 input signals from source
equipment by using dry contact switches (see figure 1-11 on page 1–21). The
monitored signals are supervised for open-circuit conditions.
Figure 1-11. Master Alarm
NOTE:
The available 60-Point Master Alarm is two 30-Point Master Alarms mounted in
one housing.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-21
Detailed Component Descriptions
Chapter 1: Introduction
1
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...