Clear Alarm Network Device List
1. To clear the alarm network device list, click on the Clear Network link. The
clear network page displays (see figure 1-44 on page 1–95).
NOTE:
This command clears the device list memory from all of the devices on the alarm
network. Immediately after the memory is cleared, the devices automatically re-
learn the device list.
Figure 1-44. Clear Network
2. Use it when you add, remove, or exchange the Master Alarm Module,
Network Communications Boards, or the Area Alarm Modules. Otherwise,
alarms will occur when the network does not report as expected.
3. To clear the alarm network’s device list, click on the Click Here link.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-95
Configuring and Browsing the Master Alarm Using the Web Pages
Chapter 1: Introduction
1
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...