Page 4-44
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Power Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.17 Power Switch
Tools required:
Needle nose pliers
5/32" hex wrench
or
Phillips head screwdriver
Removal
Shock Hazard:
Turn off the source power to the alarm, and leave it off during the installa-
tion of any component or sub-assembly. Failure to do so could result in
personal injury or equipment damage.
1. Turn off power to the alarm at the facility’s source.
2. Depending on which alarm you are replacing the power switch on, take one
of the following steps:
•
Unless the replacement is being performed on a Local Alarm, use a
5/32" hex wrench to unlock the front panel (A) (see figure 4-18 on page
4–45).
•
When changing the power supply fuse holder in the Local Alarm, use a
phillips head screwdriver to remove the front panel.
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...