Figure 1-45. Configuration Transfer
WARNING:
If one Master Alarm is configured to monitor different inputs from the
Master Alarm that you are currently configuring, ensure that the differ-
ences are accounted for in each configuration. Failure to configure the
Master Alarms correctly could endanger patient welfare.
2. Before transferring a configuration, ensure that all of the monitored signals
for both of the Master Alarms
should be
the same.
NOTE:
The configuration can be transferred to any Master Alarm on this network. This
page lists those that are available.
3. Click on the Master Alarm that you want the configuration to be transferred
to.
NOTE:
The transfer function can only be done across the Ethernet. You cannot transfer
the configuration by using the RS-232 serial port.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-97
Configuring and Browsing the Master Alarm Using the Web Pages
Chapter 1: Introduction
1
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...