Page 2-8
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Error Codes and Troubleshooting Tips
Chapter 2: Troubleshooting Procedures
Error Codes and Troubleshooting Tips
Area Alarm
The error codes for the Area Alarm Modules are used to help identify and solve
problems with the facility’s gas system or the MedPlus TotalAlert® Alarm
Network (see table 2-1 on page 2-8).
Table 2-1. Area Alarm Module Error Codes
Error Code
Error Description
Possible Corrective Action
A 01
Transducer pressure below
sensor range.
Turn on the shutoff valves, or check the
source equipment. Go to RAP 2.1 on
page 2-13.
A 02
Transducer pressure above
sensor range.
Check the source equipment, or replace
the transducer. Go to RAP 2.2 on page 2-
14.
A 03
Transducer communication
time out.
Connect the transducer, or replace the
transducer. Go to RAP 2.3 on page 2-14.
A 04
Area Alarm Module RAM
error.
Replace the Area Alarm Module. Go to
RAP 2.9 on page 2-18.
A 05
Area Alarm Module ROM
error.
Replace the Area Alarm Module. Go to
RAP 2.10 on page 2-19.
A 06
Transducer status fault.
Replace the transducer. Go to RAP 2.4 on
page 2-15.
A 07
Incorrect transducer connected
to Area Alarm Module.
Install the correct gas-specific transducer.
Go to RAP 2.5 on page 2-16.
A 08
Area Alarm Module pro-
grammed as vacuum, but units
are not inHg or mmHg.
Install the correct gas-specific Area
Alarm Module. Go to RAP 2.8 on page
2-18.
A 09
Area Alarm Module pro-
grammed as pressure, but are
units not in PSI or kPa.
Install the correct gas-specific Area
Alarm Module. Go to RAP 2.8 on page
2-18.
A 10
Transducer is programmed as
an invalid type of gas.
Install the correct gas-specific transducer.
Go to RAP 2.6 on page 2-16.
A 11
Transducer short circuit
detected.
Replace the transducer wiring. Go to
RAP 2.7 on page 2-17.
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...