2.29 No Audible Signal on the Master Alarm
1. When an alarm occurs, the audible indicator works correctly.
Yes
No
L J
Replace the Master Module (refer to procedure 4.2 on page 4-
10). If this solves the problem, go to “Final Actions” on page 2-
7. Otherwise, call BeaconMedæs Technical Support.
2. Go to “Final Actions” on page 2-7.
2.30 No Audible Signal on the Area Alarm
1. When an alarm occurs, the audible indicator works correctly.
Yes
No
L J
Replace the Area Alarm P.C. board located on the back of the
front panel (refer to procedure 4.8 on page 4-24). If this solves
the problem, go to “Final Actions” on page 2-7. Otherwise, call
BeaconMedæs Technical Support.
2.
Go to “Final Actions” on page 2-7.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 2-37
No Audible Signal on the Master Alarm
Chapter 2: Troubleshooting Procedures
2
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...