MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Local Alarm Labeling
Chapter 1: Introduction
3. Using a phillips head screwdriver, loosen the four screws (CN) enough to
slide the front panel (CO) up and off the back box (CP) (see figure 1-67 on
page 1–130).
Figure 1-67. Local Alarm Labeling
4. Slide the front panel up (CO) and off the back box (CP).
SHOCK HAZARD:
Turn off the power supply in the alarm, and leave it off during the installa-
tion of any component or sub-assembly. Failure to do so could result in
personal injury or equipment damage.
5. Turn the power switch (CQ) to the OFF position.
6. Unplug the buzzer wiring (CR).
7. Unplug the power supply wiring (CU).
8. Remove the front panel (CO).
9. For ease of handling while you are labeling; turn the front panel (CO) around
and slide it onto the screws (CN).
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...