Transducer Pressure Below Sensor Range
1. Error code
A 01
displays on an Area Alarm Module.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. The shutoff valve is open.
Yes
No
L J
Open the shutoff valve. If this solves the problem, go to “Final
Actions” on page 2-7. Otherwise, go to step 3.
3. Transducer is connected to the pipeline.
Yes
No
L J
Connect the transducer to the pipeline (refer to procedure 4.20
on page 4-50). If this solves the problem, go to “Final Actions”
on page 2-7. Otherwise, go to step 4.
4. Transducer has a slow steady flash on its heartbeat LED.
Yes
No
L J
Replace the transducer (refer to procedure 4.23 on page 4-62). If
this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 5.
5. If the problem still exists, call BeaconMedæs Technical Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 2-13
Transducer Pressure Below Sensor Range
Chapter 2: Troubleshooting Procedures
2
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...