CAUTION:
When a contractor installs this product, he is responsible for installing it in
accordance with the latest edition of the NFPA 99 standard or CSA
Z305.1-92 code dictated by the authority having jurisdiction. Any product
modifications beyond those required by the installation instructions are
the responsibility of the contractor. If these practices are not followed, the
authority having jurisdiction may not allow the facility to open.
To comply with the requirements of the Federal Communications Commission,
the following text has been included in this manual in accordance with Title 47
CFR Part 15: This equipment is designed to provide reasonable protection
against harmful interference when it is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the operation and maintenance manual,
may cause harmful interference with radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference, in which
case users will be required to correct the interference at their own expense.
To ensure compliance with the Class A Federal Communications Commission
(FCC) limits, shielded cables must be used with this unit where specified. This
equipment has been tested and found to comply with the limits for a Class A,
FCC, digital apparatus pursuant to part 15 of the rules, and with Canadian
ICES-003, Issue 3.
To comply with the requirements of the City of New York Department of
Buildings; Materials and Equipment Acceptance Division (M.E.A.), no wiring
shall enter through the knockouts in the top of the rough-in box. The wiring shall
only enter the rough-in box through the knockouts in the sides or the bottom.
When configuring the alarm, do not set the alarm return time to “never.” Set the
alarm to return every 30 minutes.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-47
Specifications
Chapter 1: Introduction
1
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...