Configuring Options Available for the Area Alarm
By using the front panel keys, you can configure or view the following options
(refer to table 1-5 on page 1-33 and table 1-6 on page 1-33).
Table 1-5. Configuring Options Available for the Area Alarm by Using the
Front Panel
Table 1-6. Viewing Options Available for the Area Alarm by Using the Front
Panel
Option
Description
dFLt
Sets alarm to default values
9ASt
Gas type
ScAL
Measuring scale
trEu
Transducer firmware revision
ArEu
Area Alarm Module firmware revision
Option
Description
A hi
Alarm high setpoint
A lo
Alarm low setpoint
A rt
Alarm repeat time
d ir
Alarm wing direction
FLor
Alarm floor level
ArEA
Alarm area
2onE
Alarm zone
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-33
Detailed Component Descriptions
Chapter 1: Introduction
1
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...