WARNING:
In operating systems, variations in pressure can cause disruption of vital
life support equipment. Do not vary pressures without the appropriate
medical facility personnel’s permission and knowledge. Varying pressure
can endanger patient welfare.
3. Before changing any pressure setpoints, notify the appropriate medical facili-
ty personnel.
4. Configure
A hi
. This is the highest that the pressure can get without
alarming.
The Area Alarm Modules are factory-programmed to NFPA 99 standard values
and with the units of measure, either metric or English, that were specified by the
customer (see table 1-4 on page 1–31).
5. Configure
A Lo
.
This is the lowest that the pressure can get without
alarming.
6. Configure
A rt
. This is the amount of time that the alarm stays silent. The
available options are as follows:
•
No
, audible alarm does not repeat.
•
30
, audible alarm repeats every 30 minutes.
•
60
, audible alarm repeats every 60 minutes.
•
90
, audible alarm repeats every 90 minutes.
•
120
, audible alarm repeats every 120 minutes.
NOTE:
To comply with the requirements of the City of New York Department of
Buildings; Materials and Equipment Acceptance Division (M.E.A.), do not set
the alarm return time to “never.” Set the alarm to return every 30 minutes.
7. Configure
d ir
. This is the direction of the wing that the module is moni-
toring. The available options are as follows:
•
____
means no direction is set.
•
nort
is north.
•
EASt
is east.
•
Sout
is south.
•
WUES
is west.
8. Configure
FLor
. This is the floor level that the module is monitoring. You
can choose from
- 2
, which are sub-basement levels, through
150
.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-119
Configuring the Area Alarm by Using the Front Panel
Chapter 1: Introduction
1
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...