a. Install the two screws (J) securing the mounting bracket (G) to the rough-
in box (K).
b. Install the two screws (F) securing the power supply assembly (C) to the
mounting bracket (G).
c. Connect the two flag terminals (D) from the power supply assembly (C)
to the Area Alarm Breakout P.C. Board (E).
11. Turn on the incoming source power to the alarm.
12. Check for gas leaks.
13. If there are gas leaks, repeat step 2 through step 6 until no gas leaks occur.
14. To ensure proper operation, perform the “Function Checks” on page 2–5.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 4-61
Local-Mount Transducer and DISS Demand Flow Valve Assembly for the Conversion Kit Housed in the
Area Line Pressure Alarm (0321) Rough-In Box
Chapter 4: Removal, Replacement, and Adjustment Procedures
4
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...