Figure 4-10. Area Alarm Module
Shock Hazard:
Turn off the power supply in the alarm, and leave it off during the installa-
tion or replacement of any component or sub-assembly. Failure to do so
could result in personal injury or equipment damage.
3. Turn the power switch (G) to the OFF position.
4. Push open the ears (E) of the connectors that hold the ribbon cable (B).
5. Remove the ribbon cable (B) from the Area Alarm Module (F).
6. Using a phillips head screwdriver, remove the two screws (C) from the retain-
er ring assembly (D and I).
7. Press the retaining clips (H), and pull the retainer ring (I) off of the Area
Alarm Module (F).
8. Remove the Area Alarm Module (F) and the retainer ring assembly
(D and I).
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 4-27
Area Alarm Module
Chapter 4: Removal, Replacement, and Adjustment Procedures
4
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...