By using the Web pages, you can configure or browse the following options
refer to table 1-2 on page 1-25 and table 1-3 on page 1-26).
Table 1-2. Configuring Options Available for the Master Alarm by Using the
Web Pages
A Combination Alarm that has a 30-Point Master Alarm with three Area Alarm Modules cannot
interface with a building automation system. The Area Alarm Modules use the same connector that
would otherwise interface with the building automation system.
Option
Description
User administrators
Sets up to three users.
User language
Sets the language for English, Spanish, or French.
Set clock
Sets the real time clock.
Device name
Sets a unique name for the Master Alarm.
Gas or vacuum type
Sets the type of gas or vacuum an input is monitoring.
Alarm message
Sets an alarm input message.
Alarm system
Sets the system number of the source.
LED mapping
Sets the source input to an LED on the front panel.
Location
Sets the location of the room where the alarm is located.
Direction
Sets the direction of the wing where the alarm is located.
Alarm return time
Sets the time between audible alarms.
Electronic notification
Sets up to five electronic notifications (combination of elec-
tronic mail and pager addresses).
Configuration transfer
Transfers the configuration of one Master Alarm to another.
Download event log
Downloads the event log to a computer file.
Download configura-
tion
Downloads the Master Alarm’s configuration to a computer
file.
Clear network memory Clears the device list from the Master Alarm’s memory.
Network setup
Sets the Ethernet network settings.
N2 address
Sets the address for the building automation system’s notifica-
tion option.
Software update
Programs the Master Alarm’s FLASH software.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-25
Detailed Component Descriptions
Chapter 1: Introduction
1
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...