Page 3-10
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Master Alarm Wiring Conventions
Chapter 3: Theory of Operation
•
For the high pressure sensor, maintain polarity by wiring the normally
closed contact wire to the positive (+) terminal and the common contact
wire to the negative (-) terminal.
• Polarity should also be observed when wiring relays. When possible, use
relays that have a separate set of contacts for each output.
Output Relays on Master Alarms
There are two dry contact relays (MAIN ALARM and AUX ALARM) on the
Master Alarm Breakout P.C. Board that are available to provide signals to out-
puts. These outputs can be wired to satisfy either normally closed or normally
open requirements from another system.
• MAIN ALARM RELAY
•
When the Master Alarm is in alarm condition, this relay activates.
•
When the Master Alarm is silenced, this relay deactivates.
•
When the Master Alarm power is off, this relay activates until the power
comes back on.
• AUX ALARM RELAY
•
When the Master Alarm is in alarm condition, this relay activates.
•
When the Master Alarm is silenced, this relay stays activated.
•
When the Master Alarm condition is remedied, this relay deactivates.
The Master Alarm also has an output relay that can be used to interface with a
building automation system.
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...