Transducer
A green LED inside the transducer’s transparent housing indicates whether the
transducer is functioning properly (see table 2-4 on page 2-11).
Table 2-4. Transducer LED Troubleshooting Codes
Error Codes
Possible Cause
Corrective Action
Slow steady flash
Transducer is functioning
properly.
No action is required.
Rapid flash
Device error.
Replace the transducer.
LED off
No power to the transducer,
or the transducer has failed.
Connect the power, or
replace the
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 2-11
Error Codes and Troubleshooting Tips
Chapter 2: Troubleshooting Procedures
2
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...