MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Medical Gas Network System Interaction
Chapter 3: Theory of Operation
Medical Gas Alarm Network System Interaction
The information and figures in this section are an overview of
typical
wiring
conventions and the interaction between components of a typical MedPlus
TotalAlert® Alarm Network (see figure 3-1 on page 3–7).
When the installation of the MedPlus TotalAlert® Alarm Network is done prop-
erly, the wiring complies with National Electrical Code® (NEC®) requirements
and the following general requirements:
• Electric service to each alarm panel must be supplied from the emergency (or
essential power) electrical power system.
• Signal wires must be a minimum of 22-gauge, stranded, shielded, twisted-
pair.
If the wire runs in or passes through an area within the ceiling or air duct desig-
nated as
Return
or
Supply Air
, plenum wire may be required. Wire in these
spaces must comply with different requirements for temperature ratings, smoke
ratings, etc.
• Signal wiring must be less than 4500\q (1372 m) in length from a signal
device to an alarm.
• Signal wiring should
not
run near any wiring carrying AC input power (typi-
cally 120V), especially in the same conduit.
• If signal wiring conduit must
cross
the AC input power wiring conduit, cross
them at a 90° angle.
• All wiring must comply with NEC® requirements.
• The MedPlus TotalAlert® Alarm Network Conversion Kits use a Class 2
power supply as the kits are listed by UL. Therefore, all sensor and switch
wiring receiving power from this system must meet the requirements as
defined in article 725 of the NEC® requirements.
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...