MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Detailed Component Descriptions
Chapter 1: Introduction
NOTE:
Adapters are needed for the following gas types: vacuum, helium, nitrogen, waste
anesthetic gas disposal (WAGD), carbon dioxide, and carbon dioxide/oxygen
mixture.
The local-mount transducer can
only
be used with the MedPlus TotalAlert®
Alarm Network Conversion Kits (see table 1-16 on page 1–43).
WARNING:
Do not remove the threaded insert from the transducer. Doing so may
result in attaching the transducer to the incorrect gas or vacuum type,
endangering patient welfare.
Never remove the threaded insert (DW) from the transducer; it is not a servicea-
ble component. If the transducer (DT) needs serviced or calibrated, send both the
transducer and the threaded insert (DX) back to the factory.
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...