Page 4-48
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Ribbon Cables
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.19 Ribbon Cables
Tools required:
Phillips head screwdriver
5/32" hex wrench
Removal
1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-20 on
page 4–48).
Figure 4-20. Ribbon Cable
Shock Hazard:
Turn off the power supply in the alarm, and leave it off during the installa-
tion of any component or sub-assembly. Failure to do so could result in
personal injury or equipment damage.
2. Turn the power switch (D) to the OFF position.
3. Push open the ears (B) of the connectors holding the ribbon cable (C) that
you want to remove.
4. Remove the ribbon cable (C).
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...