Area Alarm Labeling
Tools required:
5/32" hex wrench
11/32" nut driver
Avery® labels (Avery® part number 8167)
1. If the labeling for the Area Alarm was not completed during installation,
complete it now.
NOTE:
Each Area Alarm Module is factory-labeled with its gas or vacuum type.
2. Using a 5/32" hex wrench, unlock the front panel (CI) (see figure 1-54 on
page 1–109).
Figure 1-54. Area Alarm Labeling (Six Module Area Alarm Shown)
SHOCK HAZARD:
Turn off the power supply in the alarm, and leave it off during the installa-
tion of any component or sub-assembly. Failure to do so could result in
personal injury or equipment damage.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-109
Area Alarm Labeling
Chapter 1: Introduction
1
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...