5. Remove the source equipment wiring.
6. Push open the ears of the connectors (E) to remove the ribbon cables (B).
7. Using needle nose pliers, pull the flag terminals (F) off of the Breakout P.C.
Board (C) connectors (H).
NOTE:
There may be auxiliary wiring to the building automation system that needs to be
disconnected.
8. Pull the Breakout P.C. Board (C) off of the mounting pegs (D).
Replacement
CAUTION:
Maintain the polarity of the source equipment leads between the Master
Alarms, otherwise they may not operate as intended.
1. To replace the Breakout P.C. Board, reverse the order of the removal instruc-
tions.
2. To ensure proper operation, perform the “Function Checks” on page 2–5.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 4-15
Master Alarm Breakout P.C. Board
Chapter 4: Removal, Replacement, and Adjustment Procedures
4
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...