Master Alarm or Area Alarm Transformer
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.5
Master Alarm or Area Alarm Transformer
Tools required:
5/32" hex wrench
Phillips head screwdriver
Needle nose pliers
Removal
Shock Hazard:
Turn off the power to the alarm, and leave it off during the installation or
replacement of any component or sub-assembly. Failure to do so could
result in personal injury or equipment damage.
1. Turn off the power to the alarm at the facility’s supply.
2. Using a 5/32" hex wrench, unlock the front panel.
NOTE:
For clarity, the front panel is not shown in the illustration for this procedure.
3. Using a phillips head screwdriver, remove the five screws (A) that secure the
cover plates (B) and (C) to the rough-in assembly (D) (see figure 4-6 on page
4–17).
Содержание MedPlus TotalAlert
Страница 13: ...Page xii MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Table of Contents NOTES...
Страница 17: ...Page 1 4 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 1 Introduction NOTES...
Страница 203: ...Page 3 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 3 Theory of Operation NOTES...
Страница 289: ...Page 5 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 5 Parts List NOTES...
Страница 367: ...Page 7 2 MedPlus TotalAlert Alarm Network Operation and Maintenance Manual 205728 Chapter 7 Accessories NOTES...
Страница 370: ...Schematic Wiring Diagram Typical Local Alarm...
Страница 371: ...Schematic Wiring Diagram Typical Master Alarm...
Страница 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Страница 373: ...Schematic Wiring Diagram Typical Area Alarm...
Страница 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Страница 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...