– Set the
Data Termination
to -2V
– Set the
Alignment BER Threshold
to 1E-6.
– Set the
Clock Setup
to
Clock Data Recovery
to get the error detector's
clock from the incoming data stream.
8
Enable the pattern generator outputs by pressing the
0V (Disable)
button.
9
Press
Sync Now
and then
Auto Align
to find the optimum sampling
point. Check that the synchronization and the alignment were
successful. None of the error indicators should show red.
The resulting BER should be zero.
How to Execute the Output Levels Measurement
To run the Output Levels measurement:
1
Switch to the
Analysis
panel and then press the
Output Levels
icon.
2
Press the
Properties
button to open the
Properties
dialog box.
3
On the
Parameters
tab, set the
Sample Threshold
parameters to suitable
values that fit to the expected signal levels, for example, -1.85V to
-0.85V. Set the
Resolution
to 10 mV.
Advanced Analysis
5
Agilent J-BERT N4903 High-Performance Serial BERT
199
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...