Table 26
S.No
Parameter
Description
2
Fall time
Measures the mean
transition-time taken by
the data on the falling edge
of the eye diagram. The
data crosses three
thresholds: lower,
crossover point, upper and
the eye transition.
Note: Fall time is
dependent on the
Transition Time of the
View Tab. The two options
are 10/90 and 20/80.
3
Sample Count
The number of points
measured for the eye
diagram. If persistence,
‘infinite’, is enabled, the
value will gradually
increase.
4
0-Level
Zero level is the
measurement of the mean
value of the logical 0 of an
eye diagram.
Note: This measurement is
made in a section of the
eye referred to, as the eye
window boundaries. The
default value for the NRZ
eye window is the central
20% of the bit period. You
can change the eye
window boundary settings
in the Configure
Measurement dialog box.
These settings determine
what portion of the eye will
be measured.
Advanced Analysis
5
Agilent J-BERT N4903 High-Performance Serial BERT
273
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...