Unavailable Values
Under certain circumstances, some numerical results are not available.
This is indicated by
<invalid>
or
<not applicable>
in the numerical results
table below the measurement graphic.
•
<invalid>
indicates that the value could not be calculated. This is the
case, for example, for phase margin, when the BER threshold is set
to high values and does not intersect with the bathtub curve.
•
<not applicable>
indicates that the value could be calculated, however
is not shown because quality criteria are not met. This is the case,
for example, for RJ/DJ results, when the r^2 value of one or both
edges is <0.75. Even though RJ/DJ values can be calculated in this
case, they are not shown because confidence in the results is too low.
Explanation of the Fast Total Jitter Measurement Results
The Fast Total Jitter measurement provides both graphical and
numerical results:
The example below shows a copied result, and the display of measured
points was enabled.
The results of the recent measurement are disabled, because they refer
to a standard "bathtub" measurement that provides many more results.
By actuating the
Show
indicator, you can inspect both alternatively.
The result graph shows you the points in time that have been
investigated and whether the actual BER at these points was higher or
lower than the BER threshold specified for the measurement. The grey
area in the middle indicates the uncertainty band (see
the Fast Total Jitter Measurement” on page 178
5
Advanced Analysis
194
Agilent J-BERT N4903 High-Performance Serial BERT
Explanation of the Result Display
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...