8
Enable the pattern generator outputs by pressing the
0V (Disable)
button.
9
Press
Sync Now
and then
Auto Align
to find the optimum sampling
point. Check that the synchronization and the alignment were
successful. None of the error indicators at the top of the user
interface should show red and the resulting BER should be zero.
How to Execute the DUT Output Timing/Jitter Measurement
To run the DUT Output Timing/Jitter measurement:
1
Switch to the
Analysis
area. If the
Output Timing
screen is not yet
displayed, press the
Output Timing
icon.
2
Press the
Start
button to execute the measurement.
The measurement software runs the measurement and displays the
results. The following illustration shows the bathtub curve of the
measured bit error rate and-in the tabular view-the calculated results
for the
Optimal Sampling Delay
, the
Skew
and the
Phase Margin
.
How to Optimize the View of the Results
After you have run a measurement, the resulting graph and the
calculated numerical values are displayed. To improve the results, you
can change the measurement parameters:
1
Press the
Properties
button to open the
Properties
dialog box.
2
Use the different tabs in this dialog box to make the required settings:
Advanced Analysis
5
Agilent J-BERT N4903 High-Performance Serial BERT
183
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...