3
Open the
Edit Alias
window and change the default VISA alias name
"UsbDevice1" to "D4916".
For detailed instructions refer to the online Help.
Connecting to Pattern Generator and DUT
After the USB port has been configured, you can remove keyboard and
mouse and make the signal connections.
1
Use the N4915A-004 cable or a 2.4 mm cable (m-m) longer than 350
mm and connect the pattern generator's Data Out port to the Input
port of the N4916A.
2
Mount the SMA 50 Ohm termination on the unused Data Out port of
the pattern generator using the SMA to 2.4 mm adapter.
3
Enable the De-Emphasis Signal Converter as described in
“Understanding the Output Level Parameters” on page 73
levels such that the DUT will not be damaged.
Customizing the Instrument
9
Agilent J-BERT N4903 High-Performance Serial BERT
413