3
Click
Data Center
to move the sampling point horizontally towards
the center of the eye at the selected voltage. You may click
Cancel
at
any time.
4
Now you can use the numeric keypad or front panel knob to manually
adjust the value in the
Data Delay
field.
The
Sample Point
in the eye diagram moves horizontally as you change
the value.
T I P
While adjusting the data input delay, you can monitor the BER bar on the analyzer
display, or listen to the BER warning tones. Higher BERs may indicate that you are
measuring at the edges of the data eye.
N O T E
Pattern synchronization can be lost during sampling point adjustment. If Auto
Sync is enabled, the Serial BERT will resynchronize the patterns automatically. If
the manual synchronization mode is selected, you need to synchronize the error
detector by clicking the Pattern Sync icon and then the Sync Now button.
How to Adjust the 0/1 Decision Threshold
1
Press the
Sampling Point Setup
icon in the
ED Setup
menu.
2
Clear the
Avg. 0/1 Threshold
checkbox.
3
Click
0/1 Threshold Center
to move the sampling point vertically
towards the center of the eye at the selected delay. You may click
Cancel
at any time.
4
Now you can use the numeric keypad or front panel knob to manually
adjust the value in the
0/1 Threshold
field.
The
Sample Point
in the eye diagram moves vertically as you change
the value.
T I P
While adjusting the 0/1 threshold, you can monitor the BER bar on the analyzer
display, or listen to the BER warning tones. Higher BERs may indicate that you are
measuring at the edges of the data eye.
N O T E
Pattern synchronization can be lost during sampling point adjustment. If Auto
Sync is enabled, the Serial BERT will resynchronize the patterns automatically. If
the manual synchronization mode is selected, you need to synchronize the error
detector by clicking the Pattern Sync icon and then the Sync Now button.
4
Setting up the Error Detector
130
Agilent J-BERT N4903 High-Performance Serial BERT
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...