In this case, the signal at the Gate In port controls the timing of
synchronization and error counting for each burst.
Understanding Error Detector Output Ports
The error detector provides several output ports that are used to
connect it to other devices for further error analysis:
•
Trigger Out
Use this port to send a trigger signal to another connected device,
such as an oscilloscope.
•
Error Out
This signal can be used to trigger an external instrument to help in
error analysis. When any number of errors are located within a 128-
bit data block, a single trigger pulse is sent. If continuous errors are
detected, the error out signal would be a clock signal.
Error Out
Error
detected
Data IN
A
n Errors within
128bit granulation
(n<128)
single error
marking
A
n Errors exceeding 128bit
granulation (n>128)
•
Aux Out
This port can be used to output the clock signal or the data signal to
another connected device.
The following figure shows how the clock signal is directed to Aux
Out in Clock mode:
4
Setting up the Error Detector
112
Agilent J-BERT N4903 High-Performance Serial BERT
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...