![Agilent Technologies J-BERT N4903 Скачать руководство пользователя страница 298](http://html1.mh-extra.com/html/agilent-technologies/j-bert-n4903/j-bert-n4903_user-manual_2868694298.webp)
Such measurements can be used to find out the benchmark data, for
example, for a device prototype.
Test Methods
You can analyze the behavior of your device by employing several test
methods. If the changes in the test setup affect the BER, you can
immediately notice it on the screen. The resulting changes in the BER
let you learn about the DUT's behavior and limitations. Some examples
of modifications that you can make are listed below.
•
Add errors to the data stream. See
for details. You can, for example, perform a
confidence test of your measurement setup. The analyzer should
recognize all errors that you add. If it does not, you may have a
problem with your measurement setup. See
for details.
•
Use alternating patterns to switch between different data streams.
For example, if your device is designed to turn off in response to a
high BER, you can check if your device behaves as expected. And you
can observe what the BER is prior to data loss. See
Alternation - Concepts ” on page 102
•
Manipulate the signal being generated by the pattern generator, as
described in
“Setting up the Pattern Generator - Concepts ” on page
•
Adjust the sampling point to see the effect on the BER. See
Point Setup - Concepts ” on page 126
•
Make modifications to your device setup.
This can be used to see how changes to your device affect the BER.
You can determine what adjustments improve or degrade the BER.
Instantaneous Measurements - Procedures
This section describes how to run, monitor, and analyze instantaneous
measurements.
Running Instantaneous Measurements
The Serial BERT's error detector immediately starts calculating the bit
error rate as soon as it receives a valid signal at the data ports.
6
Evaluating Results
298
Agilent J-BERT N4903 High-Performance Serial BERT
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...