The number of errors can be set to
10
,
100
, or
1000
. When the selected
number of errors has occurred, the accumulation ends after the next
full second.
•
Number of Bits
Select this option to configure the error detector to accumulate bit
errors until the specified number of bits have been examined.
The number of bits can be set from
1E7
to
1E15
. When the selected
number of bits have been exceeded, the accumulation ends.
N O T E
The actual test period may contain more bits than specified.
N O T E
The accumulation period cannot be changed while accumulation is progress.
Browse
Click this button to browse for the directory where you wish to save the
measurement log files.
N O T E
Changes to the parameters of the accumulation setup do not take effect until the
start of the next accumulation.
BER Location
BER Location - Concepts
The Serial BERT can be run in BER location mode. In this mode, the
errors are not counted on all bits of the data signal, but on a particular
bit or block of bits in the pattern. You can specify which bit position(s)
are considered during the measurement.
This allows, for example, to calculate the BER for the header bits or the
payload of a data stream exclusively.
4
Setting up the Error Detector
154
Agilent J-BERT N4903 High-Performance Serial BERT
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...