Delete Preset
Use the
Delete Preset
button to delete the currently selected preset from
the list.
Transition Density
The
transition
density is available only with option CTR / upgrade UTR.
Number of transitions in the incoming data divided by the total number
of bits transmitted. In this field enter the transition density in (%). Click
'measure' to measure and update the transition density.
This parameter affects the loop bandwidth, and thus must be either
entered correctly or measured. Some standards specify the loop
bandwidth for a given transition density. In such a case enter the value
given in the specification, so that the CDR behaves according to the
standard. If a standard from the preset list is selected, this field is
preset.
Loop Bandwidth
The loop bandwidth is only available with option CTR / upgrade UTR.
This is the range of the CDR loop bandwidth. In this field the user should
enter the loop bandwidth value; the range is within 100 KHz to 12MHz.
If a standard from the preset list is selected, this field is preset.
When the BERT error detector is used to characterize a data stream
instead of a receiver, the loop parameters should be set according to the
used standard.
If the CDR is used to recover the bit stream from a receiver to be
characterized due to a lack of a clock output, choose the loop bandwidth
significantly higher than the receiver’s bandwidth. To characterize a
DUT’s CDR it is the best practice to use its recovered clock output
instead of the BERT’s built in CDR. Choose a low loop bandwidth to
measure the jitter on an incoming data stream as the CDR will track the
incoming jitter up to the loop bandwidth and thus make it invisible to
the error detector.
Advanced
Advanced
is available only with option CTR / upgrade UTR.
Clicking on this opens the advanced CDR setup page.
Advanced CDR setup
4
Setting up the Error Detector
124
Agilent J-BERT N4903 High-Performance Serial BERT
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...