We do not need to know the exact BER values at x+ and x-. It is sufficient
to assure that BER(x-) is greater than 10
-12
and BER(x+) is lower than
10
-12
at a confidence level of 95 %.
The algorithm then assumes that x
L
(for the left-hand slope of the
bathtub curve) is in the middle of the bracketing interval. After
repeating the procedure to determine x
R
(for the right-hand slope), it
calculates the total jitter peak-to-peak like in the standard timing/jitter
measurement.
The duration of a Fast Total Jitter measurement depends on:
•
the BER threshold
•
the bit rate
•
the sample delay step size
•
the contribution of random jitter
•
the contribution of deterministic jitter
Compared to a conservative bathtub measurement, the Fast Total Jitter
measurement can reduce the measurement time by more than a factor
of 40.
Typical test times are approximately 20 minutes at 10Gbit/s and slightly
more than one hour at 2.5 Gbit/s for a measurement at the 10
-12
BER
threshold with an overall confidence level better than 90%.
Advanced Analysis
5
Agilent J-BERT N4903 High-Performance Serial BERT
181
Measurement Duration
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...