If differential mode is selected, both input ports need to receive a
signal. The actual data signal is measured as the voltage difference
between the two incoming signals.
•
Normal
In normal mode, only the Data In port receives the data signal, the
Data In port is inactive.
•
Complement
In complement mode, only the Data In port receives the data signal,
the Data In port is inactive.
Data Inverted
Activate this checkbox to invert the polarity of the error detector
reference pattern. This function is required if your device inverts data.
Threshold
Only available, if the input is set to normal or complement mode.
Choices are:
•
Averaging
: The 0/1 threshold voltage is derived from the input signal
via a low-pass filter. This will work fine for most applications.
•
Manual input: Enter a value for applications that do not provide a
continuous data stream at the input (for example, any application
using bursts), because the averaged 0/1 threshold voltage will drift
from the correct level when there is no input. The manually set
threshold voltage must of course be within the input range.
Termination
In this field, enter the termination voltage that is appropriate for the
incoming data signal. This selection should be made before the device
is connected to the analyzer.
If the input is set to differential mode, a termination voltage can only
be set for a device that is DC-connected.
C A U T I O N
Selecting the wrong terminations may damage your device.
The Data In port is connected to a 50 Ohm load impedance (or
termination) within the error detector. Data termination refers to the
voltage level at the end of this load. The logic output from a device
Setting up the Error Detector
4
Agilent J-BERT N4903 High-Performance Serial BERT
115
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...