If the fundamental BER of your device is higher than the sync
threshold BER, you can select a higher sync threshold BER.
Error
This indicator turns red when errors are detected. View the
BER
bar or
BER Results to see the nature of the errors.
BERT connected looped back or to your device:
•
Stable errors caused by the error add function: Turn error add off.
•
Stable errors caused by false sync: Select a lower sync threshold
BER.
•
Variable and high errors may be caused by faulty connectors/cables.
BERT connected to your device:
•
Stable or variable errors can also be caused by your device. Correct
the problems with your device.
•
If the problems are not clear, then accumulate measurement data
and analyze results.
Measurement Problems
Measurement Problems - Concepts
Accumulated measurements can help you determine the cause of bit
errors. Bit errors can be caused by a variety of problems. In addition,
instrument settings can affect how errors are measured and displayed.
The sync mode setting has the greatest effect.
8
Solving Problems
378
Agilent J-BERT N4903 High-Performance Serial BERT
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...