![Agilent Technologies J-BERT N4903 Скачать руководство пользователя страница 351](http://html1.mh-extra.com/html/agilent-technologies/j-bert-n4903/j-bert-n4903_user-manual_2868694351.webp)
•
Upwards linear
starts from the minimum. A step size of 100 mUI, for
example, will result in a sequence of 0.1 UI, 0.2 UI, 0.3 UI, and so on.
The test for one frequency stops when the BER limit is exceeded or
the maximum amplitude is reached.
For logarithmic search, you can set the minimum amplitude and the
ratio. The ratio is the factor by which the present amplitude value is
multiplied to calculate the next amplitude.
•
Downwards logarithmic
starts from the maximum (which depends on
the jitter frequency). A ratio of 0.666, for example, may result in a
sequence of 1000 UI, 666 UI, 444 UI, and so on.
The test for one frequency stops when the BER limit is met or the
specified minimum amplitude is reached.
•
Upwards logarithmic
starts from the specified minimum amplitude. A
ratio of 1.5, for example, may result in a sequence of 0.1 UI, 0.15 UI,
0.225 UI, 0.3375 UI, and so on.
The test for one frequency stops when the BER limit is crossed or
the maximum amplitude (which depends on the jitter frequency) is
reached.
For binary search, you can set the accuracy.
This defines the minimum jitter amplitude step size. The test for one
frequency stops when this step size is reached.
For a description of this method refer to
For extended binary search, you can set the coarse and fine accuracy.
The test for one frequency stops when the BER limit is crossed with fine
accuracy amplitude steps.
For a description of this method refer to
.
“Optimal Measurement Setup” on page 371
Jitter Tolerance Tests
7
Agilent J-BERT N4903 High-Performance Serial BERT
351
Logarithmic search
Binary search
Extended binary search
Optimal Measurement Setup
Select Point
HTML Report
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...