A parameter that must be set for any jitter type is
Amplitude
. This
defines the timely width of the jitter component.
3
Observe the consumption of delay line capacity indicated by the
delay line markers. Ensure that the total jitter remains within the
margins of the chosen delay line.
4
Return to the overview display by pressing the
Overview
button.
5
Proceed with the next jitter type you wish to add.
N O T E
When you activate an External jitter source (by pressing its checkbox), you will find
that this apparently consumes the whole free capacity of the delay line (at the right-
hand side of the red bar).
But this is just an indication. You can still add or change internal jitter components.
The external jitter range is automatically updated when you add, change, or switch
off one of the internal sources.
Jitter Setup - Reference
The
Jitter Setup
function is used for composing the total jitter in a defined
and calibrated way.
Jitter Enable
This checkbox is used enable or disable jitter generation.
7
Jitter Tolerance Tests
328
Agilent J-BERT N4903 High-Performance Serial BERT
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...