Understanding the Result Parameters
All result parameters are calculated from the BER threshold contour
line and its bounding box. So, all parameters change with the BER
threshold.
The result display of the Eye Opening measurement shows the
maximum eye opening time, the maximum eye opening voltage, and the
position of the optimum sampling point.
These parameters are derived as illustrated in the figure below:
Error Location Capture
Error Location Capture - Concepts
The Error Location Capture measurement allows to capture the
position of an errored bit in a memory-pased pattern. The instrument
saves the position of the errored bit and writes a bit sequence
neighbouring the errored bit to a file.