CLK IN
DATA IN
AUX OUT
CDR
CLOCK MODE
Recovered CLK
Comparator
0/1 Threshold
internal
Signal
Analyzer
Data In Port Termination
To ensure a valid setup and to protect the devices from damage, proper
termination must be specified for both Data In connections. You can
specify the termination by entering the termination voltage in the
respective field.
C A U T I O N
Selecting the wrong termination may damage your device.
Why Can Wrong Terminations Damage Your Device?
Choosing wrong terminations may cause your device to output voltage
levels that are not as expected. It may also cause excessive current or
current flow in the wrong direction, which can damage your device.
N O T E
The clock termination is set to 50 Ohm, AC-coupled, and cannot be changed.
Setting up the Error Detector
4
Agilent J-BERT N4903 High-Performance Serial BERT
113