Clock
Trigger Out
Alternate Pattern Trigger Level
N O T E
This option applies for alternating patterns only.
Select this option to set the trigger level high whenever alternate pattern
B is sent.
The trigger output for 1024-bit patterns looks as follows:
Data Out
A
A
B
A
B
B
A
Trigger Out
N O T E
This pattern must be at least 1024 bits long.
Alternate Pattern Trigger Pulse
Select this option to send a trigger pulse whenever the pattern being
sent changes (A to B or vice versa). For patterns with a length divisible
by 512, there is one trigger pulse sent for every pattern repetition.
Data Out
Trigger Out
A
B
A
B
B
If the length of the pattern is not divisible by 512, a trigger is sent
whenever the pattern reaches a 512-bit RAM boundary.
Sequence Trigger
This checkbox becomes available after a user-defined sequence has been
downloaded to the pattern generator.
Click this checkbox to switch the Trigger Out port from
Clock divided by
n
to
Sequence
mode.
In
Sequence
mode, the Trigger Out can generate a spike whenever the
execution of a block starts or restarts. Whether that happens for a
Setting up the Pattern Generator
3
Agilent J-BERT N4903 High-Performance Serial BERT
95
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...