6
Press
Connect
to activate the connection between the Serial BERT
and the N4916A.
You should see the setting of the de-emphasis value and the
Disconnect
button:
If you could proceed until here, the connection of the user software to
the N4916A is ok. If the screen does not show the
De-Emphasis
entry
field, there is no connection to the N4916A.
Check the Output of the N4916A Using a Scope
1
Set the Data offset to 0 V and the amplitude to 1 V. Set the De-
Emphasis to 6 dB.
2
Connect the Output of the N4916A to an oscilloscope.
3
Connect the trigger output of the Serial BERT to the trigger input of
the oscilloscope.
Then the oscilloscope should show an output eye like the following:
If you see a display like above everything is ok.
Solving Problems
8
Agilent J-BERT N4903 High-Performance Serial BERT
389
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...