Trigger Output - Procedures
To set up the pattern generator's Trigger Output port:
1
Connect the external instrument to the Trigger Out port.
2
Click
Trigger Output
in the PG Setup menu.
3
Select the trigger pattern that you want to be generated.
“Trigger Output - Reference” on page 94
for descriptions of the
available trigger signals.
N O T E
If you have downloaded a user-defined sequence of patterns to the pattern
generator, the pattern-related settings are ignored. In this case, you can either set
up the Trigger Out port to generate a subrate of the present clock or to signal the
beginning of a sequence block (see also
“Sequence Block Parameters” on page
for details).
Trigger Output - Reference
You can always generate a divided clock signal at the Trigger Out port.
But if you wish to use the Trigger Out port for controlling external
equipment, please note that you can also generate a divided clock signal
at the pattern generator's Sub Clock Out port.
The alternate trigger signals refer to patterns and are not generated in
sequence mode.
To support the generation of a trigger spike at the beginning of a
sequence block, the Trigger Output can be put into
Sequence
mode.
Clock Divided by n
Select this option to send a trigger signal from the Trigger Out port at
every
nth
clock pulse. Note that the trigger signal itself consists of n/2
bits high followed by n/2 bits low. For example,
Clock divided by 8
works
as shown below.
3
Setting up the Pattern Generator
94
Agilent J-BERT N4903 High-Performance Serial BERT
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...