N O T E
Because the right-hand slope of the tub does not provide additional information on
the jitter, the measurement's jitter display shows only the portion at the left-hand
side of the optimum sampling point.
The jitter histogram allows you to visually inspect the jitter components:
•
Random Jitter (RJ)
•
Deterministic Jitter (DJ)
•
Estimated Total Jitter (TJ)
“Jitter Measurement Parameters” on page 192
components are calculated.
A Gaussian marker is available in this view to help analyze the jitter
components. See
“The Gaussian Marker” on page 169
for details.
Furthermore, the following parameters can be set on the
View
tab of the
Properties
dialog box:
You can analyze for:
•
All Errors
To calculate the BER values from all bits/errors.
•
Errors if 0s Expected
To calculate the BER values if "0" is expected, but "1" received.
•
Errors if 1s Expected
To calculate the BER values if "1" is expected, but "0" received.
You can calculate measurement parameters for:
•
0 Errors
To calculate the parameters for 0 errors.
Advanced Analysis
5
Agilent J-BERT N4903 High-Performance Serial BERT
189
Analyze
Calculate
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...