Timing Unit Definitions
The timing unit specifies the timebase for the measurements. It is
possible to switch between the unit interval (UI) or seconds for the time
scale. The timebase is set on the
View
page of the
Properties
dialog box.
The following figure and table shows the relationship between UI and
seconds for a 500MHz signal:
An example of the measured range of time expressed in the different
units is listed below:
Table 17
Timebase
Earliest
Sampling
Optimum
Sampling
Point
Latest
Sampling
Time
rel.
-1.5 ns
0 ns
+1.5 ns
abs.
54.2 ns
55.7 ns
57.2 ns
Unit interval
rel.
-0.75 UI
0 UI
+0.75 UI
abs.
27.1 UI
27.85 UI
28.6 UI
The Unit Interval
Unit interval values are a convenient way to express time values in a
dimensionless form. One unit interval is the equivalent of one clock
period. To convert a time value to a unit interval value, divide it by the
pulse period or multiply it by the clock frequency.
Advanced Analysis
5
Agilent J-BERT N4903 High-Performance Serial BERT
167
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...