Number of Errors
After this number of errors, the measurement stops for the current
sample point and moves to the next one. Thisallows you to speed up
the measurement. You can switch off this option if only the number
of compared bits is important.
N O T E
The measurement moves to the next sample point when the first of the two criteria
is reached.
Set the criteria for the sample threshold:
•
Resolution
Specifies the distance between the sampling points. The lower this
value is, the more sampling points you have in the selected voltage
range. The minimum step width is hardware-dependent. At the time
being, the minimum is 1 mV.
•
Low Level
This is the lower end of the measured voltage range.
•
High Level
This is the upper end of the measured voltage range.
•
Edge Resolution Optimization
Turns the resolution optimization on or off. If this option is enabled,
the Serial BERT intelligently sets a resolution so that there are more
sampling points at the edges. This can greatly improve the results
without dramatically increasing the duration of the test.
In the following example, a step width of 2 mV was used. This has
led to 502 data points per Volt.
If
Edge Resolution Optimization
is enabled, the software first scans the
threshold range in relatively coarse steps and determines the areas
where the signal produces no edges-that means, where the BER is
nearly constant.
5
Advanced Analysis
204
Agilent J-BERT N4903 High-Performance Serial BERT
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...