Running the auto-search functions and viewing eye results at
different BER thresholds will allow you to see the eye margin at
specific BERs. You can also construct basic eye contour information.
Avg. 0/1 Threshold
Activate this checkbox to make the error detector continuously set the
0/1 threshold level to the average DC voltage level of the incoming data
signal.
N O T E
Manual 0/1 threshold changes are impossible while this function is selected.
This function is useful for providing a "starting point" for sampling
somewhere within the data eye on the vertical voltage axis. It is
recommended to use this prior to clock/data alignments.
When used with the following patterns, this function may provide a 0/1
threshold that requires no further adjustment:
•
2^n-1 PRBS patterns
•
Patterns with 50% mark densities (large amplitudes and no excessive
noise)
Bit Recovery Mode
Click this button to put the error detector in bit recovery mode (BRM).
For details on BRM see
“Understanding Bit Recovery Mode” on page
Then, click
Auto Align
to set the sampling point.
Sampling Point Display
This area of the
Sampling Point Setup
window graphically illustrates the
parameters of the current sampling point and data eye. Click anywhere
in the sampling point display to set the sampling point to that location.
Setting up the Error Detector
4
Agilent J-BERT N4903 High-Performance Serial BERT
135
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...