To modify the value, click inside the text field and enter the desired
value directly with the numeric keyboard, or use the knob close to the
numeric keyboard to fine-tune the value.
The peak-to-peak amplitude is calculated by multiplying the rms value
with the crest factor (which is currently fixed to 14). The
Amplitude p-p
box shows how much of delay line capacity the setting
consumes.
The random jitter source is equipped with a 10 MHz high pass and a
500 MHz low pass filter to limit the spectral range. Both filters can be
disabled or enabled by pressing the respective buttons.
Bounded Uncorrelated Jitter Parameters
Bounded uncorrelated jitter is characterized by
•
Data rate
•
Amplitude
•
PRBS polynomial
•
Low pass filter
A graph indicates whether the chosen setting of frequency and
amplitude is tolerable.
Jitter Tolerance Tests
7
Agilent J-BERT N4903 High-Performance Serial BERT
331
Amplitude p-p
Filter